K-space exploration


The readout MR signal is stored in K-space which is equivalent to a Fourier plane. To go from a k-space data to an image requires using a 2D inverse Fourier Transform.


Frequency-encoding and phase-encoding are done so that data is spatially encoded by differences in frequency and phase, amenable to analysis by Fourier transform. In k-space, fx-coordinates (horizontal spatial frequencies) and fy-coordinates (vertical spatial frequencies) of the Fourier plane are replaced by kx and ky-coordinates.


A classic spin echo sequence fills the k-space line by line. Here is the explanation of the k-space trajectory:

  • 90° RF pulse + Slice-selection gradient : location at origin (center) of k-space
  • Negative and strong phase-encoding gradient: moves to the lower bound of k-space
  • Positive frequency-encoding gradient (dephasing lobe): moves to the right bound, location at the lower right corner
  • 180° RF pulse + Slice-selection gradient : moves to the opposite location, location at the upper left corner
  • Positive frequency-encoding gradient + Data acquisition: moves to the right + acquire MR signal
  • Repetition for each line with increasing phase-encoding gradient strength (negative to positive intensity). The amount of gradient phase change between adjacent line is constant. This results in a sequential (line by line) filling of k-space from top to bottom.


The k-space location (kx and ky coordinates) of data is governed by the accumulated effect of gradient events and excitation pulses. Here are the rules for moving in k-space:

  • The initial RF excitation pulse (with the slice-selection gradient) is the beginning of the sequence: location is at the center of k-space.
  • A 180° RF pulse causes a jump to the opposite location.
  • The greater the net strength of the phase-encoding gradient (or the longer the gradient is on), the farther from the k-space origin the data belong, in the upper direction if the gradient is positive or in the lower direction if the gradient is negative. As the duration of phase-encoding gradient is most often constant, the strength of the phase-encoding gradient governs the location on the vertical axis (ky-coordinate).
  • The longer the frequency-encoding gradient is on (or the greater the net strength of the gradient is), the farther from the k-space origin the data belong, in the right direction if the gradient is positive or in the left direction if the gradient is negative. As the strength of the frequency-encoding gradient is most often constant, the duration of the frequency-encoding gradient governs the location on the horizontal axis (kx-coordinate).