False Positive Findings for Ischemia Due to Improper Axes Alignment
Data di pubblicazione: 22 ottobre 2015 | Aggiornato il 7 gennaio 2016
Rapporto
Diagnosis
False Positive Findings for Ischemia Due to Improper Axes Alignment
History
A 60 year-old male with moderate risk factors for cardiovascular disease and an incident of chest pain presents for myocardial perfusion imaging in your Nuclear Medicine department. What is the most likely cause of the findings in the inferolateral wall?
Findings
Stress (superior rows) and rest (inferior rows) sestamibi myocardial perfusion images along with associated polar maps.
The axes are not properly aligned/not matched between stress and rest MPS slices (dashed yellow lines) causing false positive finding for ischemia in the inferolateral wall. After correction (images not displayed here), the study appeared normal.
DDx
1. Improper technical processing
2. Ischemia
3. Attenuation artifact
Discussion
It is important that images are reviewed for potential technical/processing errors such as in this case. If the interpreting physician suspects causes such as patient motion, misalignment of the axes, etc., they should ask their technologist to reprocess them. Many times, the physician has access to a dedicated workstation where he or she can make these corrections themselves. As a word of caution,be aware that if reading from a PACS system, that there may not be the capability to check for these items, and interpretations may not be properly made unless the necessary corrections are made and resent to PACS.
References
Burrell SA, Macdonald A. Artifacts and pitfalls in myocardial perfusion imaging. J Nucl Med Technol. 2006; 34:193-211.